Fast FRAP-SIA method to determine antioxidant capacity

Willmann Antonio Jiménez-Morales,María Del Pilar Cañizares-Macias,Willmann Antonio Jiménez Morales,María del Pilar Cañizares Macias
DOI: https://doi.org/10.1016/j.talanta.2024.125813
IF: 6.1
2024-03-12
Talanta
Abstract:One of the most used methods to measure antioxidant capacity in food is the ferric reducing antioxidant power (FRAP) test, which is simple, sensitive, and economical, nevertheless has long analysis times, causing measurement errors due to the instability of the FRAP reagent due to its precipitation sequential injection analysis (SIA) is a flow technique that can correct these disadvantages because it is more quickly. So, a novel FRAP-SIA method was developed to evaluate the antioxidant capacity. The system was optimized using a central composite design for hydrodynamic and chemical factors, resulting in a flow rate of 35 μL s −1 , and aspirate volumes of 33 μL–38 μL–33 μL for the sequence (FRAP-Antioxidant-FRAP). FRAP reagent was prepared with an HCl solution at 0.005 mol L −1 , improving its stability 24 times, concerning when it is in acetate buffer at pH 3.6. The method showed excellent accuracy (RSD <3%) with a LOD of 1.0 μmol L −1 of Trolox for a linear range of 5–120 μmol L −1 . The reaction time was diminished by 96% concerning the FRAP-microplate assay (from 30 min to 1.2 min). The method was applied in beverages and extracts, obtaining recovery values ranging from 91.24 to 114.22%.
chemistry, analytical
What problem does this paper attempt to address?