Multicentric characterization of organ-based tube current modulation in head computed tomography: A dosimetric and image quality study

Séléna Pondard,Corentin Desport,Mélodie Munier,Nicolas Kien,Hervé Rousseau,Olivier Merignac,Christian Popotte,Ramiro Moreno
DOI: https://doi.org/10.1016/j.ejmp.2024.103389
IF: 3.119
2024-06-01
Physica Medica
Abstract:Highlights • Dose reduction rate on the anterior part of the gantry is between 11% and 74%. • Image noise and SNR increased by up to 10% using OBTCM. • Dose is unchanged on the posterior part of the gantry on GE and Canon CT scanner. • Tube exit dose increases on the posterior part on Siemens CT scanner. Purpose To evaluate the efficiency of organ-based tube current modulation (OBTCM) in head Computed Tomography (CT) for different radiology departments and manufacturers. Materials and methods Five CT scanners from four radiology departments were evaluated in this study. All scans were performed using a standard and a routine head protocol. A scintillating fiber optic detector was placed directly on the gantry to measure the tube exit kerma. Image quality was quantified on a 16-cm HEAD phantom by measuring the signal-to-noise ratio (SNR) and the standard deviation of the Hounsfield units (HU) of circular regions of interest placed in the phantom. The Noise Power Spectrum (NPS) was also studied. Measured values were compared on images with and without OBTCM. Results The reduction rates in tube exit kerma, on the anterior part, vary between 11 % and 74 % depending on the CT scanner and the protocol used. The tube exit kerma on the posterior part remains unchanged in GE and Canon CT scanners. On the contrary, the tube exit kerma to the posterior part increases by up to 39 % in Siemens CT scanner. Image noise and SNR increase by up to 10 % in the five CT scanners. Nonetheless, the differences in noise and SNR are statistically significant (p-value < 0.05). The analysis of the NPS indicates that the noise texture remains unchanged. Conclusion OBTCM reduces the tube exit kerma to the anterior part of the gantry without reducing substantially image quality for head protocols.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?