Characterization of a Li-6 loaded liquid organic scintillator for fast neutron spectrometry and thermal neutron detection

C. D. Bass,E. J. Beise,H. Breuer,C. R. Heimbach,T. Langford,J. S. Nico
DOI: https://doi.org/10.48550/arXiv.1206.4036
2012-06-18
Instrumentation and Detectors
Abstract:The characterization of a liquid scintillator incorporating an aqueous solution of enriched lithium chloride to produce a scintillator with 0.40% Li-6 is presented, including the performance of the scintillator in terms of its optical properties and neutron response. The scintillator was incorporated into a fast neutron spectrometer, and the light output spectra from 2.5 MeV, 14.1 MeV, and Cf-252 neutrons were measured using capture-gated coincidence techniques. The spectrometer was operated without coincidence to perform thermal neutron measurements. Possible improvements in spectrometer performance are discussed.