A TEMPOL and rapamycin loaded nanofiber-covered stent favors endothelialization and mitigates neointimal hyperplasia and local inflammation

Rui Wang,Jian Lu,Jiasheng Yin,Han Chen,Hongmei Liu,Fei Xu,Tongtong Zang,Rende Xu,Chenguang Li,Yizhe Wu,Qilin Wu,Xiang Fei,Meifang Zhu,Li Shen,Junbo Ge
DOI: https://doi.org/10.1016/j.bioactmat.2022.04.033
IF: 18.9
2023-01-01
Bioactive Materials
Abstract:An increased level of reactive oxygen species (ROS) plays a major role in endothelial dysfunction and vascular smooth muscle cell (VSMC) proliferation during in-stent thrombosis and restenosis after coronary artery stenting. Herein, we report an electrospun core-shell nanofiber coloaded with 4-hydroxy-2,2,6,6-tetramethylpiperidine 1-oxyl (TEMPOL) and rapamycin (RAPA) that correspondingly serves as an ROS scavenger and VSMC inhibitor. This system has the potential to improve the biocompatibility of current drug-eluting stent (DES) coatings with the long-term and continuous release of TEMPOL and rapamycin. Moreover, the RAPA/TEMPOL-loaded membrane selectively inhibited the proliferation of VSMCs while sparing endothelial cells (ECs). This membrane demonstrated superior ROS-scavenging, anti-inflammatory and antithrombogenic effects in ECs. In addition, the membrane could maintain the contractile phenotype and mitigate platelet-derived growth factor BB (PDGF-BB)-induced proliferation of VSMCs. In vivo results further revealed that the RAPA/TEMPOL-loaded covered stents promoted rapid restoration of vascular endothelium compared with DES and persistently impeded inflammation and neointimal hyperplasia in porcine models.
engineering, biomedical,materials science, biomaterials