Successful Application of a Hemoperfusion Machine in Separated System CRRT: Evaluating Effectiveness and Safety in Critically Ill Patients With AKI in Resource‐Limited Settings

Atthaphong Phongphithakchai,Thiti Sirisuksan,Sirihatai Konwai,Suntornwit Praditau‐krit,Ussanee Boonsrirat
DOI: https://doi.org/10.1111/sdi.13210
2024-05-23
Seminars in Dialysis
Abstract:Background It is imperative to note that integrated system continuous renal replacement therapy (CRRT) necessitates a sophisticated and costly apparatus, potentially limiting its availability within resource‐limited settings. The introduction of a separated system for continuous veno‐venous hemofiltration (CVVH), characterized by uncomplicated setup procedures with a hemoperfusion machine, holds promise as a feasible alternative to CRRT for critically ill patients with acute kidney injury (AKI). Methods We aimed to retrospectively analyze the effectiveness and safety of separated CRRT applied from a hemoperfusion machine in critically ill patients with AKI during the January 2015 to December 2021 period. We also examine the in‐hospital mortality rate and multivariate logistic regression analysis to uncover the factors that affect mortality. Results We included a total of 129 critically ill patients who received separated system CRRT. The SOFA score at CRRT initiation was 12.6 ± 3.8. The fluid accumulation at the day of CRRT initiation was 3900 mL (622–8172 mL) All patients received pre‐ and postdilution CVVH. The mean prescribed CRRT dose was 22.4 ± 3.1 mL/kg/h. We found no serious complications including circuit explosion and air embolism. The in‐hospital mortality rate was 68.9%. High SOFA score and positive fluid accumulation at CRRT initiation serve as predictors of survival. Conclusions Separated system CRRT using a hemoperfusion machine provides a simplified system to operate and is proven to be effective and safe in real‐life practice, especially in resource‐limited areas.
urology & nephrology
What problem does this paper attempt to address?