A. E. Afanasiev,P. I. Skakunenko,V. I. Balykin
Abstract:A method based on measuring the shift of Ramsey spectral line in an atomic fountain in the gravitational field has been proposed to develop an atomic gravimeter involving the atomic fountain on ultracold atoms. The accuracy of the measurement of the gravitational acceleration with a fountain microwave frequency standard on Cs atoms is $$\delta g = 2 \times {{10}^{{ - 6}}}g{\text{/}}\sqrt {{{\tau }_{{\text{a}}}}} $$ . The achievable accuracy at the integration time τ a = 10 000 s is $$\delta g \approx 2 \times {{10}^{{ - 8}}}g \approx 20{\kern 1pt} $$ μGal.
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a cold - atom gravimeter based on atomic fountains and microwave transitions in order to improve the accuracy of measuring the gravitational acceleration. Specifically, the author proposes a method to achieve this goal by measuring the shift of the Ramsey spectral line of the atomic fountain in the gravitational field.
### Research Background and Problem Description
1. **Development of Quantum Sensors**
- An important direction in current sensing technology is to use quantum effects to improve the sensitivity of sensors.
- Sensors based on ultracold atomic ensembles have been widely used in fundamental research, especially in verifying general relativity, detecting gravitational waves, and searching for drifts in fundamental constants.
2. **Applications of Atomic Interferometers**
- Microwave and optical frequency standards are among the most mature quantum sensors at present. High stability is achieved by designing Ramsey atomic interferometers.
- Atomic interferometers are also used in other quantum sensors, such as atomic gravimeters/accelerometers and atomic gradiometers.
3. **Possibility of Measuring Multiple Physical Quantities**
- A natural question is: Can a single quantum sensor measure multiple physical quantities simultaneously?
- For example, some studies have attempted to design quantum accelerometers to determine accelerations along three spatial axes, and there are also methods to determine accelerations and rotation speeds on two axes in one measurement.
### Proposed Solution
In this work, the author analyzes the possibility of simultaneously measuring frequency and gravitational acceleration using an atomic fountain. The specific methods are as follows:
- **Working Principle of Atomic Fountain**
- A beam of cold atoms is formed in a magneto - optical trap and accelerated to an initial velocity \( v_0 \) and rises in the vertical direction.
- After rising to the highest point, the atoms decelerate and fall due to the gravitational acceleration \( g \). During this process, the atoms pass through the microwave field area twice (the initial section and the final section).
- The microwave frequency is selected close to the resonance frequency between the hyperfine splitting components of the atomic ground state, and the microwave intensity is adjusted to be equivalent to the action time of a π/2 pulse.
- **Shift of Ramsey Spectral Line**
- Atoms are excited from one hyperfine structure sublevel \( |g\rangle \) to another sublevel \( |e\rangle \), and the total probability has interference characteristics, depending on the phase difference between the two pulses.
- The position of the Ramsey spectral line depends on the flight time \( T \), and \( T \) in turn depends on the gravitational acceleration \( g \):
\[
T=\frac{2v_0}{g}
\]
- Therefore, the position of the Ramsey spectral line also depends on the change in the gravitational acceleration \( g \).
### Measurement Methods
1. **Signal Amplitude Variation Method**
- Select a resonance peak with a specific non - zero detuning value \( \Delta \), and measure the change in the signal amplitude on the slope of the resonance curve.
- The change in the gravitational potential will cause the shift of the Ramsey spectral line, thereby changing the signal amplitude.
2. **Resonance Peak Position Variation Method**
- Measure the change in the resonance peak position \( \delta\Delta_r \). According to the formula for the maximum sensitivity region:
\[
\Delta_r=\left(n + \frac{1}{2}\right)\frac{\pi}{\tau g}
\]
- The resonance peak frequency shift is proportional to the gravitational acceleration:
\[
\delta\Delta_r\propto\delta g
\]
### Sensitivity Estimation
- **Maximum Sensitivity**
- Under the given parameters, the maximum sensitivity occurs at a frequency detuning of about 43 Hz.
- For the cesium - atom microwave frequency standard (\( \nu = 9.2 \, \text{GHz} \)), within an integration time of 10000 seconds, the accuracy of gravitational acceleration measurement can reach \( 120 \, \mu\text{Gal} \).
### Conclusion
This method can achieve simultaneous measurement of frequency and gravitational acceleration based on a standard microwave atomic fountain. Although its accuracy is not the highest among modern gravimeters, its advantage lies in that it can be realized with existing equipment.