Protocol optimization for abdominal imaging using photon-counting CT: a consensus of two academic institutions

Fides R. Schwartz,Daniele Marin,Ludovica Lofino,Andres Abadia,Thomas O’Donnell,Bari Dane
DOI: https://doi.org/10.1007/s00261-024-04254-3
IF: 2.4
2024-03-28
Abdominal Radiology
Abstract:PurposePhoton-counting detector CT (PCD CT) is a promising technology for abdominal imaging due to its ability to provide high spatial and contrast resolution images with reduced patient radiation exposure. However, there is currently no consensus regarding the optimal imaging protocols for PCD CT. This article aims to present the PCD CT abdominal imaging protocols used by two tertiary care academic centers in the United States.MethodsA review of PCD CT abdominal imaging protocols was conducted by two abdominal radiologists at different academic institutions. Protocols were compared in terms of acquisition parameters and reconstruction settings. Both imaging centers independently selected similar protocols for PCD CT abdominal imaging, using QuantumPlus mode.ResultsThere were some differences in the use of reconstruction kernels and iterative reconstruction levels, however the individual combination at each site resulted in similar image impressions. Overall, the imaging protocols used by both centers provide high-quality images with low radiation exposure.ConclusionThese findings provide valuable insights into the development of standardized protocols for PCD CT abdominal imaging, which can help to ensure consistent as well as high-quality imaging across different institutions and allow for future multicenter research collaborations.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?