CCQM-K91.2022 key comparison on pH of an unknown phthalate buffer

Fabiano Barbieri Gonzaga,Leonardo da Silva Pardellas,Dániel Nagy,Zsófia Nagyné Szilágyi,Lyudmila Dimitrova,Ayu Hindayani,Jazmin Montero-Ruiz,José Luis Ortiz-Aparicio,Matilda Roziková,Jimena Torrez,Maria Gamboa,Ariel Galli,Angeles Rinaldi,Cintia Berón,Yuri Berta,Victoria Gelabert,Simone Fajardo,Xiu Hongyu,Wu Bing,He Xinyang,Masaki Ohata,Igor Maksimov,Emrah Uysal,Serap Gençtürk Tosun,Lokman Liv,Sergey Prokunin,Vladimir Dobrovolskiy,Darya Vengina
DOI: https://doi.org/10.1088/0026-1394/61/1a/08014
2024-10-09
Metrologia
Abstract:Main text Key comparison CCQM-K91.2022, a repetition of the previous comparison CCQM-K91 from 2011, was performed to evaluate the degree of equivalence between pH measurement results of an unknown phthalate buffer reported by participating National Metrology Institutes (NMIs) and Designated Institutes (DIs). The participants used the highest-level metrological method existing at their institution. The nominal pH value of the buffer was 4.0 at 25 °C and the suggested measurement temperatures were 5 °C, 15 °C, 25 °C, 37 °C, and 50 °C. There was a good agreement among the results from most participants at 15 °C, 25 °C, and 37 °C. Due to the lack of results, no evaluation was possible of results at 5 °C and 50 °C. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?