Performance and safety of volume guarantee ventilation in neonates using the fabian ventilator, a multicenter study

Zoltan Molnar,Ruud W. van Leuteren,Maria Wilińka,Tomasz Szczapa,Jeroen Hutten,Amit Gupta,Pawel Piwowarczyk,Katarzyna Wróblewska‐Seniuk,Karolina Chojnacka,Anton H. van Kaam,Gusztav Belteki
DOI: https://doi.org/10.1002/ppul.27264
IF: 4.0899
2024-09-14
Pediatric Pulmonology
Abstract:Objective To evaluate the performance (i.e., agreement between set and measured parameters) and safety (adverse events, device malfunctions, and ventilator alarms) of the fabian HFOi neonatal ventilator in volume guaranteed (VG) mode during conventional ventilation. To analyze the impact of leakage around the endotracheal tube and the set maximum allowed inflating pressure (Pmax). Design Prospective multicenter observational study. Methods Clinical and ventilator data were collected from 71 infants receiving VG ventilation for ≥12 h in four neonatal intensive care units (NICUs). Ventilator settings, parameters, and alarms were downloaded with 0.5 Hz sampling rate. Results Data from 4,341 h of ventilation were analyzed. The median (interquartile range, IQR) of the absolute difference between the target and measured expired tidal volume was 0.76 (0.51–1.16) mL/kg. It was less when leak was <50% (median 0.36, IQR: 0.25–0.64 mL/kg, p
pediatrics,respiratory system