Computational fluid dynamics study of respiratory mask for neonatal resuscitation

John Valerian Corda,B. Satish Shenoy,Kamarul Arifin Ahmad,Leslie Lewis,K. Prakashini,Anoop Rao,S. M. Abdul Khader,Mohammad Zuber
DOI: https://doi.org/10.1080/10255842.2024.2367120
IF: 1.6
2024-06-19
Computer Methods in Biomechanics & Biomedical Engineering
Abstract:Face cups form a vital component of breathing, assisting with devices that aid in artificial breathing for neonates. This study aims to evaluate the flow parameters in the nasal cavity for two different types of face cups. The neonatal nasal cavity model was developed from CT scans using MIMICS 21.0. Two face cups, one hemispherical and the other anatomical shaped cups are developed around the nasal cavity and the airflow is simulated using ANSYS 2021 R2. Results are compared with a nasal-only model. At the nasal valve region, the highest velocity is seen for the nasal-only model which is 16.3% higher than that of the hemispherical face cup and 15.2% superior to the anatomical-shaped face cup. In addition, the decrease in pressure across the nasal-only model is 7.4 and 6.6% below that of the hemispherical cup and anatomical cup masks. The nasal resistance values across the nasal cavity are the lowest for the nasal-only model, 7.7 and 6.7% lower respectively than the hemispherical and anatomical-shaped cups. There were very minor changes in the flow parameters such as velocity, pressure and wall shear stress when comparing the hemispherical and anatomic-shaped masks for the airflow inside the nasal cavity.
engineering, biomedical,computer science, interdisciplinary applications