Abstract:We demonstrate an arbitrary distance measurement method by chirped pulse spectrally interferometry (CPSI) using femtosecond optical frequency comb (OFC). In this paper, the chirped fiber Bragg grating (CFBG) is used to investigate the mapping relationship between displacement and the center frequency of the chirped spectral interferogram. We overcome the direction ambiguity of dispersive interferometry (DPI) ranging and expand the range of distance measurement to 18 cm. Besides, we achieve a full range of dead-zone free ranging by introducing a variable optical delay line (VODL). And through principles simulation and experiment, it is demonstrated that the measurement accuracy is 12 µm in comparison with an incremental He-Ne laser interferometer and the minimum Allen deviation is 52 nm at an average time of 1.76 ms. Similarly, in the experiment with long-distance of ∼30m, the accuracy reaches 20 µm, and 2.51 µm repeatability is achieved under harsh environment.