Simultaneous Cross-type Detection of Water Quality Indexes Via a Smartphone-App Integrated Microfluidic Paper-Based Platform.
Xiaolu Xiong,Chengwang Guo,Gengyang Yan,Bingxin Han,Zan Wu,Yueqian Chen,Shiqi Xu,Peng Shao,Hong Song,Xiyan Xu,Junfeng Han
DOI: https://doi.org/10.1021/acsomega.2c05938
IF: 4.1
2022-01-01
ACS Omega
Abstract:Water quality guarantee in remote areas necessitates the development of portable, sensitive, fast, cost-effective, and easy to-use water quality detection methods. The current work reports on a microfluidic paper-based analytical device (mu PAD) integrated with a smartphone app for the simultaneous detection of cross-type water quality parameters including pH, Cu(II), Ni(II), Fe(III), and nitrite. The shapes, baking time, amount, and ratios of reaction reagent mixtures of wax mu PAD were optimized to improve the color uniformity and intensity effectively. An easy-to-use smartphone app was established for recording, analyzing, and directly reading the colorimetric signals and target concentrations on mu PAD. The results showed that under the optimum conditions, the current analytical platform has reached the detection limits of 0.4, 1.9, 2.9, and 1.1 ppm for nitrite, Cu(II), Ni(II), and Fe(III), respectively, and the liner ranges are 2.3-90 ppm (nitrite), 3.8-400 ppm (Cu(II)), 2.9-1000 ppm (Ni(II)), 2.8-500 ppm (Fe(III)), and 5-9 (pH). The proposed portable smartphone-app integrated mu PAD detection system was successfully applied to real industrial wastewater and river water quality monitoring. The proposed method has great potential for field water quality detection.