Long–Term Astigmatism after Intraocular Pressure Sensor Implantation and Nonpenetrating Glaucoma Surgery – EYEMATE-SC Trial

Colya N. Englisch,Karl T. Boden,Peter Szurman,Kaweh Mansouri,H. Burkhard Dick,Esther M. Hoffmann,Marc J. Mackert,Paul Schütte,Achim Langenbucher,André Messias,Philip Wakili
DOI: https://doi.org/10.1097/j.jcrs.0000000000001470
2024-04-25
Journal of Cataract & Refractive Surgery
Abstract:Purpose: To investigate the long–term astigmatism after combined non-penetrating glaucoma surgery (NPGS) and implantation of the first miniaturized suprachoroidal intraocular pressure (IOP) sensor EYEMATE-SC. Setting: The study was conducted in five medical centers in two different countries. Design: Retrospective multicenter clinical study. Methods: Astigmatism of patients instrumented with the EYEMATE-SC IOP sensor was assessed over a follow-up period of three years. Refraction and corrected distance visual acuity (CDVA) were obtained preoperatively, after 6 months, 1, 2, and 3 years. A canaloplasty-operated patient cohort served as control. Astigmatism was evaluated using 3-dimensional power vector analysis involving the spherical equivalent M, and the Jackson crossed cylinder projections J 0 and J 45 . Exclusion criteria included neovascular and angle-closure glaucoma, myopia, axial length outside 22 to 26 mm, other ocular diseases, prior glaucoma surgery, other ocular surgery within 6 months (cataract surgery within 3 months) prior to NPGS, serious generalized conditions, and other active medical head/neck implants. Results: Multivariate analysis indicated no changes in astigmatism along the observation period in both the EYEMATE-SC ( n = 24) and the canaloplasty ( n = 24) group ( P > 0.05 or nonsignificant after Bonferroni correction). Astigmatism was unchanged between the EYEMATE-SC and the canaloplasty group at all time points ( P > 0.05). CDVA didn’t change along the observation period of three years in each of both groups ( P > 0.05). Conclusions: Despite its suprachoroidal localization, the present study indicates that the miniaturized EYEMATE-SC IOP sensor doesn’t negatively affect the long–term astigmatism after combined implantation with NPGS.
ophthalmology,surgery
What problem does this paper attempt to address?