Triaxial Catheter-Tip Force Sensor for MRI-Guided Cardiac Procedures

Panagiotis Polygerinos,Lakmal D. Seneviratne,Reza Razavi,Tobias Schaeffter,Kaspar Althoefer
DOI: https://doi.org/10.1109/tmech.2011.2181405
2013-02-01
IEEE/ASME Transactions on Mechatronics
Abstract:This paper presents the development and evaluation of a miniature, three-axis, fiber-optic force sensor. the sensor is manufactured using low-cost, high-resolution rapid prototyping techniques and is integrated with a catheter to enable the detection of force during MRI-guided cardiac ablation procedures. the working principle is based on reflective light-intensity modulation. A force sensitive structure (flexure) is employed to vary the distance and orientation of an integrated reflector when a force is applied at the catheter tip. in this way, the light is modulated accordingly and the force can be calculated. the sensor has a high sensitivity and an adequate linear response along all three orthogonal axes $(F_{x}, F_{y}$, and $F_{z})$ and a working range of around 0.5 N. Low-noise, high-gain electronics provide a force resolution of less than 1 gm force. Experiments demonstrate the ability of the sensor to acquire accurate force readings in a dynamic environment. MRI-compatibility experiments are performed in a clinical 1.5-T MR scanner.
What problem does this paper attempt to address?