An in-situ self-calibration method for non-contact full-field strain measurement

Huaxia Deng,Jun Wang,Haicong Zhang,Jin Zhang,Mengchao Ma,Xiang Zhong
DOI: https://doi.org/10.1016/j.measurement.2020.107871
IF: 5.6
2020-10-01
Measurement
Abstract:<p>Online calibration is important for strain measurement especially for non-contact full-field strain measurement. A self-calibration method based on modal expansion and strain–displacement relationships is developed for full-field strain measurement. The proposed method has the capability to calibrate the in–situ measured data with ground truths continuously without shutting down the test equipment. To validate the methodology for strain calibration, a simulated beam, and a cantilever beam under sinusoidal and random excitations are investigated. The post-calibrated strain of the simulated beam is better agreement with ground truths, and the <span class="math"><math>TRACε</math></span> at three points are greater than 99%. The post-calibrated strain the by videometric method, is consistent with measured data using strain gauges. And the accuracy of post-calibrated strain under sinusoidal or random excitation, is higher than that of the pre-calibrated strain. The results of experiments indicate that the proposed method is effective for calibrating strain obtained by visual measurement methods.</p>
engineering, multidisciplinary,instruments & instrumentation
What problem does this paper attempt to address?
This paper aims to solve the self - calibration problem in non - contact full - field strain measurement. Specifically, the author proposes an in - situ self - calibration method to improve the accuracy and reliability of non - contact full - field strain measurement. This method combines the finite - point displacement data and the full - field strain data and uses the modal expansion technique to optimize and calibrate the measurement results. The paper describes in detail the theoretical basis, implementation steps and experimental verification process of this method, demonstrating its effectiveness and accuracy in practical applications. ### Main problems solved in the paper: 1. **Accuracy problem of non - contact full - field strain measurement**: Traditional non - contact full - field strain measurement methods may be affected by environmental factors, equipment errors and other factors, resulting in inaccurate measurement results. The method proposed in this paper improves the accuracy of measurement results through self - calibration technology. 2. **Fusion of finite - point displacement data and full - field strain data**: How to effectively combine finite - point displacement data and full - field strain data to improve the reliability and integrity of measurement results. The method proposed in this paper realizes the effective fusion of these two kinds of data through the modal expansion technique. 3. **Optimization of the self - calibration process**: How to reduce measurement errors and improve the accuracy of measurement results through optimization algorithms (such as the least - squares method) in the self - calibration process. This paper describes the optimization process in detail and verifies its effectiveness through experiments. ### Formula examples: 1. **Strain - displacement relationship**: \[ \{\varepsilon\} = \left[\frac{\partial \Phi_n}{\partial x}\right] \left(\left[\Phi_a^T \Phi_a\right]^{-1} \Phi_a^T \{U_a\}\right) \] 2. **Optimization objective function**: \[ J(C, q) = \|\Phi q - U\|^2 + \lambda \left\|\frac{\partial \Phi}{\partial x} q - \varepsilon\right\|^2 \] 3. **Modal coordinate update**: \[ \{q\} = \left(\left[\Phi_a^T \Phi_a\right]^{-1} \Phi_a^T \{U_a\}\right) \] ### Experimental verification: The paper verifies the effectiveness of the proposed method through experiments. The experimental results show that this method can significantly improve the accuracy and reliability of non - contact full - field strain measurement under different conditions. The experimental data include time - series data of displacement and strain, as well as measurement results at different positions, demonstrating the superior performance of the method in practical applications. In conclusion, this paper proposes an innovative self - calibration method, solves the key problems in non - contact full - field strain measurement, and provides new ideas and technical support for research and application in related fields.