An Intraoperative $β^-$ Detecting Probe For Radio-Guided Surgery in Tumour Resection

Andrea Russomando,Fabio Bellini,Valerio Bocci,Giacomo Chiodi,Francesco Collamati,Erika De Lucia,Raffaella Donnarumma,Riccardo Faccini,Carlo Mancini Terracciano,Michela Marafini,Riccardo Paramatti,Vincenzo Patera,%Davide Pinci,Luigi Recchia,Alessio Sarti,Adalberto Sciubba,Elena Solfaroli Camillocci,Cecilia Voena,Silvio Morganti
DOI: https://doi.org/10.48550/arXiv.1511.02059
IF: 4.506
2015-11-06
Medical Physics
Abstract:The development of the $\beta^-$ based radio-guided surgery aims to extend the technique to those tumours where surgery is the only possible treatment and the assessment of the resection would most profit from the low background around the lesion, as for brain tumours. Feasibility studies on meningioma, glioma, and neuroendocrine tumors already estimated the potentiality of this new treatment. To validate the technique, prototypes of the intraoperative probe required by the technique to detect $\beta^-$ radiation have been developed. This paper discusses the design details of the device and the tests performed in laboratory. In such tests particular care has to be taken to reproduce the surgical field conditions. The innovative technique to produce specific phantoms and the dedicated testing protocols is described in detail.