Systematic study on the influence of inductive chargers on active personal dosemeters

Steffen Ketelhut,Hayo Zutz,Oliver Hupe
DOI: https://doi.org/10.1088/1361-6498/ad8fb7
2024-11-09
Journal of Radiological Protection
Abstract:Electromagnetic compatibility testing plays an important role in the type testing of radiation protection dosemeters in view of technical developments and the associated increase in electromagnetic fields. Lately, the use of inductive charging devices has grown as a user-friendly type of charging mobile-phones. In this article, we investigate their impact on active personal dosemeters. The measurements show a substantial additional dose reading of up to several tens of mSv when exposed to the field of an inductive charger for 20 s. According to the widely used Qi standard, the charging devices operate at frequencies of between 87 kHz and 205 kHz for power transfers between 5 W and 30 W. These parameters fall outside the scope of type-testing standards for active personal dosemeters. An update of These standards might therefore be necessary.
environmental sciences,public, environmental & occupational health,radiology, nuclear medicine & medical imaging,nuclear science & technology
What problem does this paper attempt to address?