A duplex droplet digital PCR assay for Salmonella and Shigella and its application in diarrheal and non-diarrheal samples

Jingyun Zhang,Yuanming Huang,Panpan Xue,Zhifei Zhan,Zheng Huang,Jie Li,Baowei Diao,Biao Kan
DOI: https://doi.org/10.1016/j.ijid.2022.04.040
IF: 12.073
2022-04-01
International Journal of Infectious Diseases
Abstract:Objectives: To evaluate a duplex droplet digital PCR (ddPCR) assay targeting Salmonella fimY and Shigella ipaH genes.Methods: The linear range, precision, analytical sensitivity, and analytical specificity of the ddPCR assay were analyzed. The ddPCR assay was compared with quantitative real-time PCR (qPCR) using 362 stool samples from 187 mild diarrheal and 175 non-diarrheal children.Results: The duplex ddPCR assay showed good linearity in the range of 5.3 × 100 to 1.24 × 105 copies/reaction for Salmonella and 1.9 × 100 to 1.84 × 105 copies/reaction for Shigella. When analyzed with spiked stool samples, the limit of detection and limit of quantification were 550 and 5,500 CFU/mL stool sample for Shigella, respectively; while both were 1.0 × 104 CFU/mL stool sample for Salmonella. Among 362 stool samples, more samples were detected as positive by ddPCR than by qPCR. Salmonella load was significantly higher in the diarrheal samples than in the non-diarrheal samples. Determined by receiver-operating characteristic analysis, the optimal cut-off value was 1.25 × 104 copies/mL stool sample to distinguish between symptomatic and asymptomatic Salmonella infections.Conclusions: Salmonella and Shigella prevalence may have been underestimated in the past.