Characterization of R-134a superheated droplet detector for neutron detection
Prasanna Kumar Mondal,Rupa Sarkar,Barun Kumar Chatterjee
DOI: https://doi.org/10.1016/j.apradiso.2014.02.024
IF: 1.787
2014-08-01
Applied Radiation and Isotopes
Abstract:R-134a (C2H2F4) is a low cost, easily available and chlorine free refrigerant, which in its superheated state can be used as an efficient neutron detector. Due to its high solubility in water the R-134a based superheated droplet detectors (SDD) are usually very unstable unless the detector is fabricated using a suitable additive, which stabilizes the detector. The SDD is known to have superheated droplets distributed in a short-lived and in a relatively long-lived metastable states. We have studied the detector response to neutrons using a (241)AmBe neutron source and obtained the temperature variation of the nucleation parameters and the interstate kinetics of these droplets using a two-state model.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear