New device, 'old' allergens. Allergic contact dermatitis caused by the Dexcom G7 glucose sensor

Martin Mowitz,Sarah Hosseini,Ingrid Siemund,Josefin Ulriksdotter,Cecilia Svedman
DOI: https://doi.org/10.1111/cod.14514
2024-02-06
Contact Dermatitis
Abstract:Two children with contact dermatitis caused by the Dexcom G7 (DG7) glucose sensor were patch tested with a medical device series and with acetone extracts made from different parts of the DG7 sensor. Both patients tested positive to isobornyl acrylate (IBOA), hydroabietyl alcohol, and glyceryl hydrogenated rosinate. Gas chromatography–mass spectrometry analyses of the DG7 extracts showed the presence of IBOA and colophonium‐related substances. Background Allergic contact dermatitis (ACD) has been reported as an adverse effect from the use of several glucose sensors and insulin pumps from different manufacturers. Isobornyl acrylate (IBOA) has been identified as a major culprit sensitizer, but also other acrylates and (modified) colophonium have been reported as causes of ACD. Objectives To report the two first cases diagnosed with ACD caused by the Dexcom G7 (DG7) glucose sensor. Patients and Methods Two children with suspected ACD from DG7 were patch tested with our medical device series with an addition of selected test preparations including two variants of modified colophonium – methyl hydrogenated rosinate (MHR) and glyceryl hydrogenated rosinate (GHR). Both patients were also tested with acetone extracts made from different parts of the DG7 sensor. The extracts were analysed by gas chromatography–mass spectrometry (GC–MS). Results Both patients tested positive to IBOA, hydroabietyl alcohol and GHR. In addition, patient 1 had a positive reaction to MHR and patient 2 had a positive reaction to colophonium. The GC–MS analyses showed the presence of IBOA and colophonium‐related substances in the DG7 extracts. Conclusions Both patients were diagnosed with contact allergy to well‐known medical device‐related sensitizers. The presence of IBOA and (modified) colophonium in a newly introduced (on the Swedish market in 2023) glucose sensor is remarkable and indicates an inadequate toxicological assessment of the materials used in the sensor.
dermatology,allergy
What problem does this paper attempt to address?