Evaluating the Accuracy of a Novel Motion Tracking Sensor Technology as part of a Full-Body Immersive VR Rehabilitation System

Michale Casale
DOI: https://doi.org/10.1016/j.apmr.2024.02.635
IF: 4.06
2024-04-04
Archives of Physical Medicine and Rehabilitation
Abstract:Research Objectives To evaluate the accuracy and usefulness of a novel motion tracking technology currently used as part of a commercially available, full body immersive rehabilitation system. Specifically, we wanted to assess the accuracy of different, clinically relevant kinematic measurements relevant to physical therapy clinical practice (e.g., joint angles, functional reach measurement, etc.). Design Motion tracking data was collected using sensors worn by participants guided by a protocol to produce specific movements and poses while also capturing ground truth measurements. The accuracy of the measurements produced by the motion tracking sensors against the ground truth measurements was compared in order to better understand their utility in clinical practice. Participants performed a series of scripted, clinically relevant upper body and lower body movements and poses while the (1) system performed ambient data capture and measurement and (2) a clinically-trained experimenter collected ground truth data for later evaluation. Setting Data was captured in a non-clinical environment. Participants 20 healthy participants volunteered to take part in the research. Interventions No clinical interventions were administered. Main Outcome Measures Joint angles (degrees), spatial positions (inches). Results The accuracy of 10 different measurement types produced by the system: 8 joint angles along with hand and feet positions. Most joint angles were found to be within 15 degrees of accuracy of the ground truth measurements (interquartile range 2-14 degrees) while positional accuracy of the hands had mean errors of 1.3 and 1.5 inches, respectively. Conclusions The novel motion tracking sensor technology used in conjunction with a commercially available immersive rehabilitation system was shown to have sufficient accuracy for use in many rehabilitation settings, especially as compared to the known variability associated with human measurement often used in traditional rehabilitation practice. This opens the possibility for improvements to current clinical measurements, as the sensor technology is readily available for use in clinical settings. Taken in concert, these technologies can offer new insights into a patient's health, allowing clinicians to provide better targeted therapies for patients. Author(s) Disclosures All authors listed are employees of Penumbra, Inc.
rehabilitation,sport sciences
What problem does this paper attempt to address?