Magnetic Compression Anastomosis Gastro-Jejunostomy: Feasibility and Efficacy of a Novel Device in a Swine Model

Michel Gagner,Todd Krinke,Maxime Lapointe-Gagner,J.N. Buchwald
DOI: https://doi.org/10.1016/j.soard.2024.04.451
IF: 3.709
2024-05-22
Surgery for Obesity and Related Diseases
Abstract:Background Means of addressing technical challenges in forming gastrojejunostomy (GJ) anastomoses and maintaining their patency are sought. Objectives Evaluation of preclinical feasibility and healing efficacy of a novel linear magnetic compression anastomosis (MCA) device to form a patent GJ versus sutured jejunal enterotomy (JE) sites in swine. Setting Single-center veterinary testing facility. Methods Feasibility of 3 prototype sizes (4, 6, and 8 cm) of a metal MCA device (MCAD) to form a patent GJ were evaluated over 6 weeks. A distal magnet was laparoscopically inserted in the jejunum, a proximal magnet was placed gastroscopically in the stomach; magnets were aligned to gradually form an anastomosis, self-detach, and be expelled. At necropsy, magnetic compression anastomoses were assessed for patency and compared with JE tissues to evaluate wound healing. Results MCADs aligned at the GJ location without complications. In 5/6 MCAD pairs, dislodgement occurred between 7–26 days; expulsion 13–31 days; 1 MCAD pair was retained in the stomach. At necropsy, all pigs were healthy, gaining a mean 15.0 kg. Anastomoses were not adequately patent in 2/4 pigs receiving the 4-cm or 6-cm MCADs because their linear length was too small. But, anastomoses of both pigs receiving the 8-cm MCADs maintained full patency. Minimal inflammation and fibrosis were seen in MCA specimens versus sutured enterotomies. Conclusions A novel linear magnetic compression anastomosis device was feasible and effectively created a patent gastrojejunostomy anastomosis in swine with minimal inflammation and fibrosis. The MCAD may be appropriate for clinical evaluation.
surgery
What problem does this paper attempt to address?