US extended-field-of-view imaging technology.

L Weng,A P Tirumalai,C M Lowery,L F Nock,D E Gustafson,P L Von Behren,J H Kim
DOI: https://doi.org/10.1148/radiology.203.3.9169720
IF: 19.7
1997-06-01
Radiology
Abstract:PURPOSE: To develop an ultrasound (US) extended-field-of-view scanning technique that combines the convenience of a real-time scanner with the spatial advantages of a static B-mode scanner and provides a panoramic image in real time without position sensors or cumbersome articulated arms.MATERIALS AND METHODS: An image-registration-based position-sensing technique was used to track probe motion and reconstruct a large composite image during real-time scanning. The probe motion (translation and rotation) was estimated by combining multiple local motion vectors. This computationally intensive process required a special programmable image processor.RESULTS: Large, resolution-preserved composite images up to 60 cm long were obtained. Measurement accuracy as determined with phantom experiments was better than 5%. The method could be applied to any probe or image format.CONCLUSION: In addition to providing a panoramic image to expand diagnostic capabilities, extended-field-of-view US provides a more easily interpretable image and is an effective cross-specialty communication tool.
radiology, nuclear medicine & medical imaging