Possibility of using Parallel beam CT in dental treatment by correction of movement errors with virtual Alignment Method

Seokhwan Yoon,Seong-Kyun Kim,Kyungtaek Jun
DOI: https://doi.org/10.48550/arXiv.2005.03565
IF: 4.506
2020-05-07
Medical Physics
Abstract:The purpose of this study is to correct the movement errors that may occur because of the patients' movement during the CT scanning and thus, to obtain more precise CT images, additionally to investigate the possible use of Parallel Beam CT in the dental treatment. An extracted human tooth was scanned with PBCT and the possible movement errors that patients may cause during the CT scan were intentionally assigned. The maximum width of errors was designed to reach 0.1 mm. The trajectory of the fixed point was traced in the sinogram that includes the fixed point. The part deviated from the sinusoidal curve was modified using the Virtual Alignment Method. The quality of reconstructed image was remarkably degraded when the movement errors were given. A radiopaque part inside the tooth was used as a fixed point. The errors were corrected by shifting the fixed point to the center of sinogram and this relocation of fixed point did not cause any distortion of image or change of size. The movement errors were corrected with the fixed point inside the tooth. The more precise CT images obtained through this method are expected to be of extensive application in the area of dental treatment using CAD/CAM.
What problem does this paper attempt to address?