The Adaptive Gain Integrating Pixel Detector at the European XFEL
Aschkan Allahgholi,Julian Becker,Annette Delfs,Roberto Dinapoli,Peter Goettlicher,Dominic Greiffenberg,Beat Henrich,Helmut Hirsemann,Manuela Kuhn,Robert Klanner,Alexander Klyuev,Hans Krueger,Sabine Lange,Torsten Laurus,Alessandro Marras,Davide Mezza,Aldo Mozzanica,Magdalena Niemann,Jennifer Poehlsen,Joern Schwandt,Igor Sheviakov,Xintian Shi,Sergej Smoljanin,Lothar Steffen,Jolanta Sztuk-Dambietz,Ulrich Trunk,Qingqing Xia,Mourad Zeribi,Jiaguo Zhang,Manfred Zimmer,Bernd Schmitt,Heinz Graafsma
DOI: https://doi.org/10.1107/s1600577518016077
IF: 2.557
2019-01-01
Journal of Synchrotron Radiation
Abstract:The Adaptive Gain Integrating Pixel Detector (AGIPD) is an X-ray imager, custom designed for the European X-ray Free-Electron Laser (XFEL). It is a fast, low-noise integrating detector, with an adaptive gain amplifier per pixel. This has an equivalent noise of less than 1 keV when detecting single photons and, when switched into another gain state, a dynamic range of more than 10 4 photons of 12 keV. In burst mode the system is able to store 352 images while running at up to 6.5 MHz, which is compatible with the 4.5 MHz frame rate at the European XFEL. The AGIPD system was installed and commissioned in August 2017, and successfully used for the first experiments at the Single Particles, Clusters and Biomolecules (SPB) experimental station at the European XFEL since September 2017. This paper describes the principal components and performance parameters of the system.
optics,physics, applied,instruments & instrumentation