A flow-batch lab-in-syringe foam microextraction platform for the simultaneous preconcentration and in situ membraneless gas-liquid separation of mercury prior to cold vapor atomic absorption spectrometry

Natalia Manousi,Aristidis N. Anthemidis
DOI: https://doi.org/10.1016/j.aca.2024.342208
IF: 6.911
2024-01-07
Analytica Chimica Acta
Abstract:Herein, the proof-of-concept of a novel lab-in-syringe (LIS) foam microextraction platform is presented as a front-end to cold vapor atomic absorption spectrometry (CVAAS) for the simultaneous preconcentration and membraneless gas-liquid separation (GLS) of inorganic mercury in biological samples. The proposed method is based on the on-line formation of the ammonium pyrrolidine dithiocarbamate complex with mercury that was retained in the pores of polyurethane foam (PUF) immobilized on the piston of the LIS system. Metal complex elution and in situ mercury vapor generation are accomplished inside the microsyringe in a flow-batch format, while the separation of vapor species is achieved via the membraneless GLS found at the top of the syringe's barrel. Under optimized operation conditions, for 90 s preconcentration time, the limit of detection was 0.02 μg L −1 and the repeatability (RSD) was 3.8% (at the 0.5 μg L −1 concentration level), within a working range extending up to 4.0 μg L −1 . The practicality of the novel manifold was demonstrated using the Blue Analytical Grade Index, while the accuracy of the method was evaluated using certified reference materials and spiked samples.
chemistry, analytical
What problem does this paper attempt to address?