Practical Application of Multivendor MRI‐Based R2* Mapping for Liver Iron Quantification at 1.5 T and 3.0 T

Gregory Simchick,Ruiyang Zhao,Qing Yuan,Mounes Aliyari Ghasabeh,Stefan Ruschke,Dimitrios C. Karampinos,David T. Harris,Raphael do Vale Souza,Ryan J. Mattison,Michael R. Jeng,Ivan Pedrosa,Ihab R. Kamel,Shreyas Vasanawala,Takeshi Yokoo,Scott B. Reeder,Diego Hernando
DOI: https://doi.org/10.1002/jmri.29401
IF: 4.4
2024-04-27
Journal of Magnetic Resonance Imaging
Abstract:Background Recent multicenter, multivendor MRI‐based R2* vs. liver iron concentration (LIC) calibrations (i.e., MCMV calibrations) may facilitate broad clinical dissemination of R2*‐based LIC quantification. However, these calibrations are based on a centralized offline R2* reconstruction, and their applicability with vendor‐provided R2* maps is unclear. Purpose To determine R2* ranges of agreement between the centralized and three MRI vendors' R2* reconstructions. Study Type Prospective. Subjects Two hundred and seven subjects (mean age 37.6 ± 19.6 years; 117 male) with known or suspected iron overload from four academic medical centers. Field Strength/Sequence Standardized multiecho spoiled gradient echo sequence at 1.5 T and 3.0 T for R2* mapping and a multiple spin‐echo sequence at 1.5 T for LIC quantification. MRI vendors: GE Healthcare, Philips Healthcare, and Siemens Healthineers. Assessment R2* maps were generated using both the centralized and vendor reconstructions, and ranges of agreement were determined. R2*‐LIC linear calibrations were determined for each site, field strength, and reconstruction and compared with the MCMV calibrations. Statistical Tests Bland–Altman analysis to determine ranges of agreement. Linear regression, analysis of covariance F tests, and Tukey's multiple comparison testing to assess reproducibility of calibrations across sites and vendors. A P value <0.05 was considered significant. Results The upper limits of R2* ranges of agreement were approximately 500, 375, and 330 s−1 for GE, Philips, and Siemens reconstructions, respectively, at 1.5 T and approximately 700 and 800 s−1 for GE and Philips, respectively, at 3.0 T. Within the R2* ranges of agreement, vendor R2*‐LIC calibrations demonstrated high reproducibility (no significant differences between slopes or intercepts; P ≥ 0.06) and agreed with the MCMV calibrations (overlapping 95% confidence intervals). Data Conclusion Based on the determined upper limits, R2* measurements obtained from vendor‐provided R2* maps may be reliably and practically used to quantify LIC less than approximately 8–13 mg/g using the MCMV calibrations and similar acquisition parameters as this study. Evidence Level 1 Technical Efficacy Stage 3
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?