Highly sensitive electrochemical biosensor for MUC1 detection based on a DNA functionalized CdTe quantum dots as signal enhancers

Bao-Kang Jin,Jiajia Wang,Chun Kan
DOI: https://doi.org/10.1039/d4ay01544d
IF: 3.1
2024-10-07
Analytical Methods
Abstract:In this paper, a proportional electrochemical biosensor based on cadmium telluride/ polypyrrole (CdTe/PPy) nanocomposite is developed for the detection of MUC1 with high selectivity and sensitivity. The result indicates that CdTe/PPy nanocomposite is modified on the surface of glassy carbon electrode (GCE), which supplies a large surface area for immobilizing cap-DNA, ensuring its high selectivity and sensitivity. Whereafter, CdTe linked sig-DNA (MUC1 aptamer) is introduced, allowing MUC1 aptamer to hybridize with cap- DNA. CdTe as signal amplification elements in order to generate the differential pulse voltammetry (DPV) signal. Conceivably, the target MUC1 detection was based upon the current signal change due to the quantities change of the signal amplification elements CdTe. In the presence of MUC1, the MUC1 aptamer specifically binds to MUC1, resulting in the release of CdTe-sig-DNA from the electrode surface and the decrease of peak current. Under the optimized experimental conditions, the electrochemical biosensor present a highly selective, sensitive, stability and reproducibility for the MUC1 from the range of 0.1 nM to 100 nM with detection limit of 0.05 nM (S/N = 3). Therefore, the electrochemical biosensor has a potential application in medical disease diagnosis application detection.
spectroscopy,chemistry, analytical,food science & technology
What problem does this paper attempt to address?