First in‐human evaluation of [1‐13C]pyruvate in D2O for hyperpolarized MRI of the brain: A safety and feasibility study
Kofi Deh,Guannan Zhang,Angela Hijin Park,Charles H. Cunningham,Nadia D. Bragagnolo,Serge Lyashchenko,Shake Ahmmed,Avigdor Leftin,Elizabeth Coffee,Hedvig Hricak,Vesselin Miloushev,Marius Mayerhoefer,Kayvan R. Keshari
DOI: https://doi.org/10.1002/mrm.30002
2024-01-13
Magnetic Resonance in Medicine
Abstract:Purpose To investigate the safety and value of hyperpolarized (HP) MRI of [1‐13C]pyruvate in healthy volunteers using deuterium oxide (D2O) as a solvent. Methods Healthy volunteers (n = 5), were injected with HP [1‐13C]pyruvate dissolved in D2O and imaged with a metabolite‐specific 3D dual‐echo dynamic EPI sequence at 3T at one site (Site 1). Volunteers were monitored following the procedure to assess safety. Image characteristics, including SNR, were compared to data acquired in a separate cohort using water as a solvent (n = 5) at another site (Site 2). The apparent spin–lattice relaxation time (T1) of [1‐13C]pyruvate was determined both in vitro and in vivo from a mono‐exponential fit to the image intensity at each time point of our dynamic data. Results All volunteers completed the study safely and reported no adverse effects. The use of D2O increased the T1 of [1‐13C]pyruvate from 66.5 ± 1.6 s to 92.1 ± 5.1 s in vitro, which resulted in an increase in signal by a factor of 1.46 ± 0.03 at the time of injection (90 s after dissolution). The use of D2O also increased the apparent relaxation time of [1‐13C]pyruvate by a factor of 1.4 ± 0.2 in vivo. After adjusting for inter‐site SNR differences, the use of D2O was shown to increase image SNR by a factor of 2.6 ± 0.2 in humans. Conclusions HP [1‐13C]pyruvate in D2O is safe for human imaging and provides an increase in T1 and SNR that may improve image quality.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?