A Novel Incident Circular Polarized Light Raman Optical Activity (ICP‐ROA) Spectrometer With Advanced Polarization Control

Klaus Hofmann,Luisa Martin,Ingo Fischer
DOI: https://doi.org/10.1002/jrs.6733
IF: 2.727
2024-09-19
Journal of Raman Spectroscopy
Abstract:The design of a backscatter Raman optical activity (ROA) spectrometer with incident light circular polarization (ICP) is presented. ICP is controlled using a combination of waveplates, compensating for unwanted birefringence and beam offsets. Realignment of the spectra in post‐processing reduced artifacts caused by spatial offsets. The setup was validated by measuring references samples, which show very good agreement with previously npublished results. The design and setup of a novel and simple backscatter Raman optical activity (ROA) spectrometer with incident light circular polarization (ICP) is presented, constructed from commercially available components. Incident light polarization is controlled using a combination of waveplates, compensating for unwanted birefringence and beam offsets. Realignment of the spectra in post‐processing reduced artifacts caused by spatial offsets. Spurious signals from achiral solvents like toluene and water are almost completely removed. The setup was validated by measuring references samples, including α‐pinene, carvone, and glucose in aqueous solution. The spectra show very good agreement with previously published results.
spectroscopy
What problem does this paper attempt to address?