Holography-guided procedural planning for modifying Venus P-valve implantation technique in patients with left pulmonary artery stents: a case-series

Angelo Fabio d'Aiello,Laura Schianchi,Francesca Bevilacqua,Paolo Ferrero,Angelo Micheletti,Diana Gabriela Negura,Giulia Pasqualin,Massimo Chessa
DOI: https://doi.org/10.3389/fcvm.2024.1378924
IF: 3.6
2024-05-14
Frontiers in Cardiovascular Medicine
Abstract:Background: Venus P -valveTM (Venus Medtech, Hangzhou, China) is a self-expandable bioprosthetic valve that can be transcatheter-implanted in native right ventricular outflow tract (RVOT) patients. Valve implantation is technically challenging. Due to the implantation technique, left pulmonary artery (LPA) stents represent a relative contraindication to Venus P -valve. In this case series, we describe our experience in implanting Venus P -valve in patients with previous LPA stents and the use of holographic models to facilitate procedural planning. Methods and results: From January to October 2023, 17 patients were scheduled for Venus P -Valve implantation. 16/17 (94%) patients were successfully implanted. 3/16 (18.7%) patients underwent Venus P -valve implantation with LPA stents. All patients underwent pre-operative CT scan. CT data set were employed to create three-dimensional (3D) holographic models (Artiness, Milan, Italy) of the entire heart, which were useful to plan valve implantation with a modified technique. Procedural success rate was 100%. No procedural complications occurred. All three patients presented good haemodynamic and angiographic results at discharge and follow-up visits. Conclusion: This case-series underscores the feasibility of Venus P -valve implantation in patients with previous LPA stents. The use of holographic models facilitated procedural planning in these challenging anatomical scenarios.
cardiac & cardiovascular systems
What problem does this paper attempt to address?