Optimization of a Scintillation Detector for Positron Emission Tomography

M. V. Belov,M. V. Zavertyaev,V. A. Kozlov,V. S. Tskhay
DOI: https://doi.org/10.3103/s1068335624600645
2024-08-31
Bulletin of the Lebedev Physics Institute
Abstract:We report simulation and resolution evaluation of monolithic position sensitive detectors with different surfaces. The best result was obtained using a combination of rough polishing of the crystal surface and a Teflon wrapper with a resolution in the interaction depth of dZ = 0.89 ± 0.01 mm. The best planar resolution dX = 0.53 ± 0.01 mm is obtained for a finely polished plate with an ESR film. The diffuse reflector also demonstrates an increase in light collection by 33% compared to the specular one.
physics, multidisciplinary
What problem does this paper attempt to address?