Point-of-Care Fluorescence Biosensing System for Rapid Multi-Allergen Screening

Demuru,Chai-Gao,Shynkarenko,Hermann,Boia,Cristofolini,Petkus,Generelli,Paoletti,Cattaneo,Burr
DOI: https://doi.org/10.3390/s24113280
IF: 3.9
2024-05-22
Sensors
Abstract:With the steady increase in allergy prevalence worldwide, there is a strong need for novel diagnostic tools for precise, fast, and less invasive testing methods. Herein, a miniatured fluorescence-based biosensing system is developed for the rapid and quantitative detection of allergen-specific immunoglobulin-E. An antibody-based fluorescence assay in a microfluidic-patterned slide, combined with a custom-made portable fluorescence reader for image acquisition and user-friendly software for the data analysis, enables obtaining results for multiple allergens in just ~1 h with only 80 μL of blood serum. The multiplexed detection of common birch, timothy grass, cat epithelia, house dust mite, and dog epithelia shows quantitative IgE-mediated allergic responses to specific allergens in control serum samples with known total IgE concentration. The responses are verified with different control tests and measurements with a commercial fluorescence reader. These results open the door to point-of-care allergy screening for early diagnosis and broader access and for large-scale research in allergies.
engineering, electrical & electronic,instruments & instrumentation,chemistry, analytical
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a portable fluorescent biosensing system for rapid, multi - allergen screening. With the continuous increase in the global prevalence of allergies, there is an increasingly urgent need for accurate, rapid and less - invasive detection methods. Traditional methods such as skin prick test (SPT) and oral food challenge, although fundamental in identifying and confirming allergen sensitivity, are often criticized for their invasiveness and the discomfort they cause to patients, and these techniques may not fully meet the needs of modern data - driven allergy research and personalized medicine qualitatively. For this reason, researchers have developed a portable fluorescent biosensing system based on a microfluidic chip. This system can rapidly and quantitatively detect multiple allergen - specific immunoglobulin E (IgE) within one hour using only 80 microliters of serum samples. By combining antibody fluorescence assay, a customized portable fluorescence reader and user - friendly data analysis software, this system can perform rapid allergen - specific IgE screening in various settings such as doctor's offices, pharmacies or university laboratories, thereby promoting early diagnosis, broader access to testing and large - scale allergy research. Specifically, this system can detect common allergens such as birch pollen, timothy grass, cat dander, dust mites and dog dander, and the accuracy of its detection results has been verified by comparison with commercial fluorescence readers. This not only improves the patient experience, but also makes it possible to collect detailed large - scale data, which helps to deepen the understanding of allergic diseases.