Automatic adjustment of oxygen concentration during high‐flow nasal cannula treatment using a targeted SpO2 feedback system

Woo Jung Seo,Eun Young Kim,Ga Jin Seo,Hee Jung Suh,Jin Won Huh,Sang‐Bum Hong,Younsuck Koh,Chae‐Man Lim
DOI: https://doi.org/10.1111/nicc.13033
2024-02-08
Nursing in Critical Care
Abstract:Aim To test whether targeted SpO2 feedback (TSF), an automatic control system for fraction of inspired oxygen (FiO2), achieves more time in the optimal SpO2 range and/or reduces the frequency of manual adjustments to administered FiO2 compared with conventional manual titration in patients with hypoxia on high‐flow nasal cannula (HFNC) therapy. Study Design Twenty‐two patients were recruited from two hospitals. For each, two sessions of manual mode and two sessions of TSF were applied in a random order, each session lasting 2 h. The target SpO2 on TSF was 95%. Oxygen monitoring levels were classified into four SpO2 ranges: hypoxia (≤ 89%), borderline (90%–93%), optimal (94%–96%) and hyperoxia (≥ 97%). The two modes were compared based on the proportion of time spent in each SpO2 range and the number of manual FiO2 adjustments. Results The proportion of time in the optimal SpO2 range was 20.5% under manual titration mode and 65.4% under TSF (p
nursing
What problem does this paper attempt to address?