Lesion metrics and 12‐month outcomes of very‐high power short duration radiofrequency ablation (90W/4 s) under mild conscious sedation

Peter Calvert,Ioanna Koniari,Mark T. Mills,Reza Ashrafi,Richard Snowdon,Dhiraj Gupta,Vishal Luther
DOI: https://doi.org/10.1111/jce.16269
IF: 2.7
2024-04-05
Journal of Cardiovascular Electrophysiology
Abstract:Introduction Pulmonary vein isolation (PVI) is often performed under general anaesthesia (GA) or deep sedation. Anaesthetic availability is limited in many centers, and deep sedation is prohibited in some countries without anaesthetic support. Very high‐power short duration (vHPSD—90W/4 s) PVI using the Q‐Dot catheter is generally well tolerated under mild conscious sedation (MCS) though an understanding of catheter stability and long‐term effectiveness is lacking. We analyzed lesion metrics and 12‐month freedom from atrial arrythmia with this approach. Methods Our approach to radiofrequency (RF) PVI under MCS is standardized and includes a single catheter approach with a steerable sheath. We identified patients undergoing Q‐Dot RF PVI between March 2021 and December 2022 in our center, comparing those undergoing vHPSD ablation under MCS (90W/MCS) against those undergoing 50 W ablation under GA (50 W/GA) up to 12 months of follow‐up. Data were extracted from clinical records and the CARTO system. Results Eighty‐three patients met our inclusion criteria (51 90W/MCS; 32 50 W/GA). Despite shorter ablation times (353 vs. 886 s; p
cardiac & cardiovascular systems
What problem does this paper attempt to address?