Adaptive threshold algorithm for detecting EEG-interburst intervals in extremely preterm neonates

Johannes Caspar Mader,Manfred Hartmann,Anastasia Dressler,Lisa Oberdorfer,Zsofia Rona,Sarah Glatter,Christine Czaba-Hnizdo,Johannes Herta,Tilmann Kluge,Tobias Werther,Angelika Berger,Johannes Koren,Katrin Klebermass-Schrehof,Vito Giordano
DOI: https://doi.org/10.1088/1361-6579/ad7c05
2024-09-19
Physiological Measurement
Abstract:Objective: This study provides an adaptive threshold algorithm for burst detection in electroencephalograms (EEG) of preterm infantes and evaluates its performance using clinical real-world EEG data. Approach: We developed an adaptive threshold algorithm for burst detection in EEG recordings from preterm infants. To assess its applicability in the real-world, we tested the algorithm on a dataset of 30 clinical EEG recordings which were not preselected for good quality, to ensure a real-world scenario. Main results: Interrater agreement was substantial at a kappa of 0.73 (0.68 - 0.79 inter-quantile range). The performance of the algorithm showed a similar agreement with one clinical expert of 0.73 (0.67 - 0.76) and a sensitivity and specificity of 0.90 (0.82 – 0.94) and 0.95 (0.93 – 0.97), respectively. Significance: The adaptive threshold algorithm demonstrated robust performance in detecting burst patterns in clinical EEG data from preterm infants, highlighting its practical utility. The fine-tuned algorithm achieved similar performance to human raters. The algorithm proves to be a valuable tool for automated burst detection in the EEG of preterm infants.
engineering, biomedical,biophysics,physiology
What problem does this paper attempt to address?