Primary experiment for photoacoustic human in vivo tissue strain imaging

Naoto Sato,C. Sumi
Abstract:Sumi et al have been developing ultrasonic (US) echo human tissue displacement vector measurement methods such as the multidimensional cross-spectrum phase gradient method (CSPGM) [1], the multidimensional autocorrelation method (AM) [2], etc. The methods have also been used for measuring and imaging of strain tensors and shear moduli of in vivo breast, liver, heart etc [3]. In ref. 4, the strain measurements were also performed both for photoacoustic (PA) and rf-echo signals obtained from a human in vivo wrist (our first trial [5]). A real-time PA and rf-echo imaging instrument (AcousticX, CYBERDYNE Inc., Ibaraki, Japan) was used. The layered structures of skin, vein, soft tissue and artery were successfully imaged by both axial strain imagings using the 2-dimensional (2D) CSPGM having a substantial robustness with respect to a signal noise [2]. The PA signals yielded higher axial resolution and higher contrast strain images than rf-echo signals. And, the lateral resolutions and lateral contrasts were also examined for the corresponding PA and echo images. It was confirmed that the PA signal generations depend on tissue types such as a skin and a blood. In this report, for the strain measurement, the 2D AM with a fewer calculations but with less robustness for a signal noise than the 2D CSPGM [2] is used. And, the spatial resolutions and contrats are more specifically examined for PA and echo images, i.e., including axial ones.