Automated software-assisted diagnosis of esophageal squamous cell neoplasia using high-resolution microendoscopy

Mimi C Tan,Sheena Bhushan,Timothy Quang,Richard Schwarz,Kalpesh H Patel,Xinying Yu,Zhengqi Li,Guiqi Wang,Fan Zhang,Xueshan Wang,Hong Xu,Rebecca R Richards-Kortum,Sharmila Anandasabapathy,Mimi C. Tan,Kalpesh H. Patel,Rebecca R. Richards-Kortum
DOI: https://doi.org/10.1016/j.gie.2020.07.007
IF: 10.396
2021-04-01
Gastrointestinal Endoscopy
Abstract:<h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Background and Aims</h3><p>High-resolution microendoscopy (HRME) is an optical biopsy technology that provides subcellular imaging of esophageal mucosa but requires expert interpretation of these histopathology-like images. We compared endoscopists with an automated software algorithm in esophageal squamous cell neoplasia (ESCN) detection and evaluated the endoscopists' accuracy with and without input from the software algorithm.</p><h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Methods</h3><p>Thirteen endoscopists (6 experts, 7 novices) were trained and tested on 218 post-hoc HRME images from 130 consecutive patients undergoing ESCN screening/surveillance. The automated software algorithm interpreted all images as neoplastic (high-grade dysplasia, ESCN) or non-neoplastic. All endoscopists provided their interpretation (neoplastic or non-neoplastic) and confidence level (high or low) without and with knowledge of the software overlay highlighting abnormal nuclei and software interpretation. The criterion standard was histopathology consensus diagnosis by 2 pathologists.</p><h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Results</h3><p>The endoscopists had a higher mean sensitivity (84.3%, standard deviation [SD] 8.0% vs 76.3%, p=0.004), lower specificity (75.0%, SD 5.2% vs 85.3%, p&lt;0.001) but no significant difference in accuracy (81.1%, SD 5.2% vs 79.4%, p=0.26) in ESCN detection compared with the automated software algorithm. With knowledge of the software algorithm, the endoscopists significantly increased their specificity (75.0% to 80.1%, p=0.002) but not sensitivity (84.3% to 84.8%, p=0.75) or accuracy (81.1% to 83.1%, p=0.13). The increase in specificity was among novices (p=0.008) but not experts (p=0.11).</p><h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Conclusions</h3><p>The software algorithm had lower sensitivity but higher specificity for ESCN detection than endoscopists. Using computer-assisted diagnosis, the endoscopists maintained high sensitivity while increasing their specificity and accuracy compared with their initial diagnosis. Automated HRME interpretation would facilitate widespread usage in resource-poor areas where this portable, low-cost technology is needed.</p>
gastroenterology & hepatology
What problem does this paper attempt to address?