Approaching an optimum time resolution for synchroscan streak-camera measurements with visible synchrotron light

G. Schiwietz,M. Koopmans,Ch. Kalus,M. Marongiu,A. Schälicke,P. Schnizer,M. Ries
DOI: https://doi.org/10.1016/j.nima.2024.169196
2024-05-01
Abstract:The optical beam diagnostics at the BESSY II light source in Berlin have been improved significantly over the last few years. In particular, the streak-camera system has been extended in precision and sensitivity to allow two-dimensional imaging in time and space for equilibrated and non-equilibrated bunch patterns. In this paper, we prove experimentally and theoretically that we have reached a sub-ps RMS total time resolution using filtered synchrotron light. Detailed simulations, including the different physical time-dispersion mechanisms, show the influence of various band-pass and edge wavelength filters on the resolution. The limits for unfiltered near-visible synchrotron radiation (white-light) and the band-pass filter to achieve optimal time resolution are derived as well, providing a basis for more advanced beam-dynamics studies in the near future.
physics, particles & fields, nuclear,nuclear science & technology,instruments & instrumentation
What problem does this paper attempt to address?