Development of aptamer-based lateral flow devices for rapid detection of SARS-CoV-2 S protein and uncertainty assessment

Jiachen Wang,Lianhui Zhao,Xiaotong Li,Yunhua Gao,Wei Yong,Yong Jin,Yiyang Dong
DOI: https://doi.org/10.1016/j.talanta.2024.126825
IF: 6.1
2024-09-14
Talanta
Abstract:The outbreak and spread of COVID-19 have highlighted the urgent need for early diagnosis of SARS-CoV-2. Nucleic acid testing as an authoritative tool, is cumbersome, time-consuming, and easy to cross-infect, while the available antibody self-testing kits are deficient in sensitivity and stability. In this study, we developed competitive aptamer-based lateral flow devices (Apt-LFDs) for the quantitative detection of SARS-CoV-2 spike (S) protein. Molecular docking simulation was used to analyze the active binding sites of the aptamer to S protein, guiding complementary DNA (cDNA) design. Then a highly efficient freezing strategy was applied for the conjugation of gold nanoparticles (AuNPs) and DNA probes. Under optimal conditions, the linear range of the constructed Apt-LFDs was 0.1–1 μg/mL, and the limit of detection (LOD) was 51.81 ng/mL. The cross-reactivity test and stability test of the Apt-LFDs showed good specificity and reliability. The Apt-LFDs had recoveries ranging from 89.45 % to 117.12 % in pharyngeal swabs. Notably, the uncertainty of the analytical result was evaluated using a "bottom-up" approach. At a 95 % confidence level, the uncertainty report of ( 453.37±54.86 ) ng/mL with k = 2 was yielded. Overall, this study provides an important reference for the convenient and reliable detection of virus proteins based on LFDs.
chemistry, analytical
What problem does this paper attempt to address?