Colorimetric immunoassay for rapid detection of Staphylococcus aureus based on etching-enhanced peroxidase-like catalytic activity of gold nanoparticles

Shuo Yao,Juan Li,Bo Pang,Xuechen Wang,Yujie Shi,Xiuling Song,Kun Xu,Juan Wang,Chao Zhao
DOI: https://doi.org/10.1007/s00604-020-04473-7
IF: 5.7
2020-08-19
Microchimica Acta
Abstract:A novel colorimetric immunoassay for the detection of <i>Staphylococcus aureus</i> (<i>S. aureus</i>) based on a combination of immunomagnetic separation and signal amplification via etching-enhanced peroxidase-like catalytic activity of gold nanoparticles (AuNPs) was developed. Nanoconjugates composed of gold and iron oxide nanoparticles were synthesized and further modified with anti<i>S. aureus</i> immunoglobulin Y (IgY), which was used for the selective enrichment and rapid separation of target bacteria in complex matrices. AuNPs functionalized with anti<i>S. aureus</i> aptamer were used as an artificial enzyme which has peroxidase-like catalysis activity. Catalytic activity of AuNPs is inhibited by modifying aptamer. However, catalysis of modified AuNPs remarkably enhanced by hydrogen peroxide etching. Based on collecting unbound modified AuNPs in the supernatant and 3,3′,5,5′-tetramethylbenzidine-hydrogen peroxide reporting system, the yellow color of solution decreases linearly with increasing the concentration of <i>S. aureus</i> ranging from 10 to 10<sup>6</sup> cfu/mL. The limit of detection is 10 cfu/mL, and total detection time is 65 min. The recoveries of the <i>S. aureus</i> spiked in food samples are 88.2–119.8%.
chemistry, analytical
What problem does this paper attempt to address?