PDS-MAR: a fine-grained projection-domain segmentation-based metal artifact reduction method for intraoperative CBCT images with guidewires

Tianling Lyu,Zhan Wu,Gege Ma,Chen Jiang,Xinyun Zhong,Yan Xi,Yang Chen,Wentao Zhu
DOI: https://doi.org/10.1088/1361-6560/ad00fc
IF: 3.5
2023-10-08
Physics in Medicine and Biology
Abstract:Objective: Since the invention of modern CT systems, metal artifacts have been a persistent problem. Due to increased scattering, amplified noise, and limited-angle projection data collection, it is more difficult to suppress metal artifacts in cone-beam CT, limiting its use in human- and robot-assisted spine surgeries where metallic guidewires and screws are commonly used. Approach: To solve this problem, we present a fine-grained projection-domain segmentation-based MAR method termed PDS-MAR, in which metal traces are augmented and segmented in the projection domain before being inpainted using triangular interpolation. In addition, a metal reconstruction phase is proposed to restore metal areas in the image domain. } Main results: The proposed method is tested on both digital phantom data and real scanned CBCT data. It achieves much-improved quantitative results in both metal segmentation and artifact reduction in our phantom study. The results on real scanned data also show the superiority of this method. Significance: The concept of projection-domain metal segmentation would advance MAR techniques in CBCT and has the potential to push forward the use of intraoperative CBCT in human-handed and robotic-assisted minimal invasive spine surgeries.
engineering, biomedical,radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?