A switching lung mechanics model for detection of expiratory flow limitation

Carlotta Hennigs,Franziska Bilda,Jan Graßhoff,Stephan Walterspacher,Philipp Rostalski
DOI: https://doi.org/10.1515/auto-2023-0206
2024-05-08
at - Automatisierungstechnik
Abstract:Expiratory flow limitation (EFL) is an often unrecognized clinical condition with a multitude of negative implications. A mathematical EFL model is proposed to detect flow limitations automatically. The EFL model is a switching one-compartment lung mechanics model with a volume-dependent airway resistance to simulate the dynamic behavior during expiration. The EFL detection is based on a breath-by-breath model parameter identification and validated on clinical data of mechanically ventilated patients. In the severe flow limitation group 93.9 % ± 5 % and in the no limitation group 10.2 % ± 13.7 % of the breaths are detected as EFL. Based on the high detection rate of EFL, these results support the usefulness of the EFL detection. It is a first step toward an automated detection of EFL in clinical applications and may help to reduce underdiagnosis of EFL.
automation & control systems
What problem does this paper attempt to address?