Effects of Probe Laser Intensity on Co-Magnetometer Operated in Spin-Exchange Relaxation-Free Regime

Yang Fu,Wenfeng Fan,Jiasen Ruan,Ye Liu,Zhenglong Lu,Wei Quan
DOI: https://doi.org/10.1109/tim.2022.3144738
IF: 5.6
2022-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:The probe laser intensity working point optimization of the co- magnetometer operated in the spin-exchange relaxation-free (SERF) regime is studied theoretically and experimentally in this article. Based on the steady-state solution of the Bloch equation, the error equation of output signal considering the probe laser pumping effect is established. A method for real-time measuring the pumping rate of the probe laser on SERF co- magnetometer is presented. Combined with the measurement results, a method for quickly finding the optimal working point for the probe laser intensity is presented. The signal noise and long-term stability of the co- magnetometer under different probe laser intensities are measured experimentally to verify the accuracy of the law deduced from the theory. The results show that the pumping effect of the probe laser significantly reduces the signal-to-noise ratio of the co- magnetometer, which verifies the necessity of supplementing to error equation in this article. In addition, selecting an appropriate probe laser intensity working point can effectively reduce signal noise and bias instability of the SERF co- magnetometer.
engineering, electrical & electronic,instruments & instrumentation
What problem does this paper attempt to address?