Feasibility Of Amulet Occluder Implantation After Failed Left Atrial Appendage Occlusion Attempt: Insights From The Emerge Laa Post-Approval Study

Akash Makkar,Mohamad Alkhouli,Christopher R. Ellis,Atman P. Shah,Megan Coylewright,James V. Freeman,Jordan A. Anderson,Ryan Gage,Dhanunjaya Lakkireddy
DOI: https://doi.org/10.1016/j.hrthm.2024.05.004
IF: 6.779
2024-05-20
Heart Rhythm
Abstract:Background While expertise in left atrial appendage occlusion (LAAO) has grown, certain intricate anatomies may pose challenges, rendering them unsuitable for LAAO using the selected device. Objective To characterize outcomes of patients with prior failed percutaneous LAAO procedures who underwent subsequent attempt with an Amulet occluder in the EMERGE LAA post-approval study. Methods Patients enrolled in the NCDR LAAO Registry and had an Amulet occluder implant attempt between FDA approval (August 14, 2021) and June 30, 2023 were evaluated. A safety endpoint through 7 days or hospital discharge (whichever was later) and major adverse events (MAEs) through 45 days were reported. Results A total of 8,591 patients underwent attempted Amulet occluder implantation of which 244 patients had Prior Failed LAAO. Implant success was 88.9% and 96.2% in patients with Prior Failed LAAO and Index LAAO, respectively (p 0.05). At 45 days, peri-device leak ≤3 mm was achieved in over 90% of patients in either group. Conclusion A high degree of implant success with a low rate of adverse events can be achieved using the Amulet occluder. The findings imply that the dual-occlusive mechanism Amulet occluder facilitates successful closure, even in challenging anatomical scenarios.
cardiac & cardiovascular systems
What problem does this paper attempt to address?