Optimal configurations of an electromagnetic tracking system for 3D ultrasound imaging of pediatric hips – A phantom study

Thanh-Tu Pham,Lawrence H. Le,John Andersen,Edmond H. Lou
DOI: https://doi.org/10.1016/j.medengphy.2024.104221
IF: 2.356
2024-08-26
Medical Engineering & Physics
Abstract:Tracking the position and orientation of a two-dimensional (2D) ultrasound scanner to reconstruct a 3D volume is common, and its accuracy is important. In this study, a specific miniaturized electromagnetic (EM) tracking system was selected and integrated with a 2D ultrasound scanner, which was aimed to capture hip displacement in children with cerebral palsy. The objective of this study was to determine the optimum configuration, including the distance between the EM source and sensor, to provide maximum accuracy. The scanning volume was aimed to be 320 mm × 320 mm × 76 mm. The accuracy of the EM tracking was evaluated by comparing its tracking with those from a motion capture camera system. A static experiment showed that a warm-up time of 20 min was needed. The EM system provided the highest precision of 0.07 mm and 0.01° when the distance between the EM source and sensor was 0.65 m. Within the testing volume, the maximum position and rotational errors were 2.31 mm and 1.48°, respectively. The maximum error of measuring hip displacement on the 3D hip phantom study was 4 %. Based on the test results, the tested EM system was suitable for 3D ultrasound imaging of pediatric hips to assess hip displacement when optimal configuration was used.
engineering, biomedical