Clinical Outcomes for Adult Single-Sided Deafness Cochlear Implantees Exceeding the 5% Candidacy Criterion

Joshua G. W. Bernstein,Elicia M. Pillion,Anthony M. Tolisano
DOI: https://doi.org/10.1097/aud.0000000000001578
2024-10-19
Ear & Hearing
Abstract:While single-sided deafness cochlear implants (SSD-CIs) have now received regulatory approval in the United States, candidate-ear candidacy criteria (no better than 5% word-recognition score) are stricter than for traditional CI candidates (50 to 60% speech recognition, best-aided condition). SSD implantation in our center began before regulatory approval, using a criterion derived from traditional candidacy: 50% consonant-nucleus-consonant (CNC) word-identification score in the candidate ear. A retrospective analysis investigated whether SSD patients exceeding the 5% CNC criterion nevertheless benefitted from a CI as assessed by spatial-hearing tests (speech understanding in noise [SIN] and localization) and by a patient-reported outcome measure quality-of-life instrument validated for patients with CIs.
otorhinolaryngology,audiology & speech-language pathology
What problem does this paper attempt to address?