A Self-Driven Microfluidic Chip for Ricin and Abrin Detection

Xuexin Bai,Chenyi Hu,Liang Chen,Jing Wang,Yanwei Li,Wei Wan,Zhiying Jin,Yue Li,Wenwen Xin,Lin Kang,Han Jin,Hao Yang,Jinglin Wang,Shan Gao
DOI: https://doi.org/10.3390/s22093461
IF: 3.9
2022-05-03
Sensors
Abstract:Ricin and abrin are phytotoxins that can be easily used as biowarfare and bioterrorism agents. Therefore, developing a rapid detection method for both toxins is of great significance in the field of biosecurity. In this study, a novel nanoforest silicon microstructure was prepared by the micro-electro-mechanical systems (MEMS) technique; particularly, a novel microfluidic sensor chip with a capillary self-driven function and large surface area was designed. Through binding with the double antibodies sandwich immunoassay, the proposed sensor chip is confirmed to be a candidate for sensing the aforementioned toxins. Compared with conventional immunochromatographic test strips, the proposed sensor demonstrates significantly enhanced sensitivity (≤10 pg/mL for both toxins) and high specificity against the interference derived from juice or milk, while maintaining good linearity in the range of 10–6250 pg/mL. Owing to the silicon nanoforest microstructure and improved homogeneity of the color signal, short detection time (within 15 min) is evidenced for the sensor chip, which would be helpful for the rapid tracking of ricin and abrin for the field of biosecurity.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation