Rapid detection of four pathogens in bloodstream infection by antimicrobial peptide capture combined with multiplex PCR and capillary electrophoresis

Xingxing Liu,Xue Du,Yanzhi Huang,Bo Pang,Mengfan Zhang,Yingwei Ma,Xiaomu Wang,Xiuling Song,Jinhua Li,Juan Li
DOI: https://doi.org/10.1016/j.microc.2022.108199
IF: 5.304
2022-11-18
Microchemical Journal
Abstract:Bloodstream infection (BSI) is a common and serious infection that threatens the life of patients. In this study, a rapid method combining magnetic separation techniques and multiplex PCR-based capillary electrophoresis (mPCR-CE) was developed to detect four pathogens causing BSI. Taking advantage of the binding capacity with the cell membrane, the antimicrobial peptides melittin was connected to magnetic beads as a broad-spectrum bacterial capture element, which was used to capture pathogenic bacteria in whole blood. The detection limit of this proposed method was significantly lower than that of the common PCR method, reaching 10 1 CFU/ mL with excellent specificity. The overall clinical sensitivity was 90%, and the clinical specificity was 100%. There were no false positives for testing another 100 clinical samples suspected of bloodstream infections. The detection platform took only 3-4 hours, which is obviously shorter than the identification time of the traditional methods. Our method possesses tremendous clinical application potential for its advantages in detecting mixed bacterial infections.
chemistry, analytical
What problem does this paper attempt to address?