Feasibility and usability of three consecutive nights with self‐applied, home polysomnography

Ferretti Dimitri,Anna S. Islind,Kristín A. Ólafsdóttir,Sigridur Sigurdardottir,Kamilla R. Jóhannsdóttir,Jan Hedner,Ludger Grote,Erna Sif Arnardottir
DOI: https://doi.org/10.1111/jsr.14286
2024-07-26
Journal of Sleep Research
Abstract:Summary In‐laboratory polysomnography, the gold‐standard for diagnosing sleep disorders, is resource‐demanding and not conducive to multiple night evaluations. Ambulatory polysomnography, especially when self‐applied, could be a viable alternative. This study aimed to assess the feasibility and reliability of self‐applied polysomnography over three consecutive nights in untrained participants, assessing: technical success rate; comparing sleep diagnostic variables from single and multiple nights; and evaluating participants' subjective experience. Data were collected from 78 participants (55.1% females) invited to test a self‐applicable polysomnography device for three consecutive nights at home. The technical success rate for valid sleep recordings was 82.5% out of 234 planned study nights, with 87.2% of participants obtaining at least two valid nights. Misclassification of obstructive sleep apnea severity was higher in participants with mild OSA (21.4%) compared with those with moderate‐to‐severe obstructive sleep apnea or no obstructive sleep apnea. Sleep efficiency and wake after sleep onset showed improvement from Night 1 to Night 3 (p
neurosciences,clinical neurology
What problem does this paper attempt to address?